GMP80P measurement performance
Property Description/Value
Measurement range 0–20 %CO2
Accuracy 1
At 5 %CO2 ±0.1 %CO2
At 0–8 %CO2 ±0.2 %CO2
At 8–20 %CO2 ±0.4 %CO2
Calibration uncertainty
At 5 %CO2 ±0.07 %CO2
At 20 %CO2 ±0.27 %CO2
Long-term stability
At 0–8 %CO2 ±0.3 %CO2/year
At 8–12 %CO2 ±0.5 %CO2/year
At 12–20 %CO2 ±1.0 %CO2/year
Start-up, warm-up, and response time
Start-up time at +25 °C (+77 °F) < 10 s
Warm-up time to full specification < 4 min
Response time (T90) < 1 min
Other
Sampling frequency 1 s
GMP80P operating environment
Property Description/Value

Operating environment

Indoor use

Operating temperature

+15 … +40 °C (+59 … +104 °F)

Storage temperature

−20 … +60 °C (−4 … +140 °F)

Operating humidity

0–95 %RH

Storage humidity

20–85 %RH

IP rating

IP4X
Flow rate of sampled/aspirated gas 0.3 l/min
GMP80P powering
Property Description/Value
Operating voltage 12–30 V DC
Power consumption

Typical: 0.7 W

Maximum: 0.9 W

Current consumption 60 mA typical at minimum voltage
GMP80P digital communication
Property Description/Value
Digital output

RS-485

GMP80P mechanical specifications
Property Description/Value
Cable connector type M12 5-pin A-coded male
Barbed fitting for gas outlet (optional) For tube with ID 4 mm (approx. 5/32 in)
Dimensions (H × W × D)
Probe body (incl. cable connectors), without sampling pipe: 114 × 102 × 40 mm (4.49 × 4.02 × 1.57 in)
Depth with sampling pipe: 145 mm (5.71 in)
Weight 410 g (14.46 oz)
Materials
GMP80P (excluding GMP251 probe):

Housing

Sampling pipe

Barbed fitting for gas outlet

Connectors (on the bottom)

Aluminum and polyamide

Stainless steel

Nickel-plated brass

Nickel-plated brass

GMP251 probe:

Housing

Filter

Connector

PBT polymer

PTFE membrane, PBT polymer grid

Nickel-plated brass

GMP80P compliance
Property Description/Value

EU directives and regulations

EMC Directive (2014/30/EU)

RoHS Directive (2011/65/EU) as amended by 2015/863

REACH Regulation (EC 1907/2006)

Electromagnetic compatibility (EMC)

EN 61326‑1, basic electromagnetic environment

FCC part 15 B, Class B

ICES‑3 / NMB‑3 (Class B)

Compliance marks

CE, China RoHS, RCM
GMP80P manufacturer contact information
Property Description/Value
Made in Finland by Vaisala Oyj, Vanha Nurmijärventie 21, FI-01670, Vantaa
UK address Vaisala Ltd, 6230 Bishops Court, Solihull Parkway, Birmingham Business Park, Birmingham B37 7YB
1 At 25 °C (77 °F) and 1013 hPa (incl. repeatability and non-linearity).