Example of configuring filtering factor with Indigo80 handheld indicator - GMP251 - GMP80P

GMP251 and GMP80P User Guide

Document code
M211799EN
Revision
K
Language
English
Product
GMP251
GMP80P
Document type
User guide
  • Probe connection cable

The filtering factor allows averaging the output if the measuring environment produces occasional exceptionally high or low readings.

This example shows how to configure the filtering factor as 90 using the Indigo80 handheld indicator.

This means that the new measurement output is a combination of the latest measurement (90 %) and the previous measurement (10 %).

  1. Connect the probe(s) to the indicator.
  2. Open the indicator main menu by pressing .
  3. Select Devices. If you have more than one device connected to the indicator, make a further selection between the devices.
  4. Select Settings > Measurement.
  5. Select Filtering factor.
    1. Use the arrow buttons to set the value as 90.
    2. Select Set.
  6. Exit the menu by pressing or return to the main menu by pressing .