Chemical tolerance - GMP251 - GMP252 - GMP80P

GMP251 and GMP80P User Guide

Document code
M211799EN
Revision
K
Language
English
Product
GMP251
GMP80P
Document type
User guide
Detach the GMP251 probe from GMP80P before cleaning GMP251 with hydrogen peroxide, acetone, or acetic acid.

The following chemicals can be used to clean the GMP251 probe:

  • H2O2 (2000 ppm), non-condensing
  • Alcohol-based cleaning agents such as ethanol and IPA (70 % isopropyl alcohol, 30 % water)
  • Acetone
  • Acetic acid

 

The following chemicals can be used to clean the GMP80P body:
  • Alcohol-based cleaning agents such as ethanol and IPA (70 % isopropyl alcohol, 30 % water).
Avoid exposing the probe to chemicals for unnecessarily long periods of time. Do not immerse the probe in a chemical or spray anything directly on the probe. Wipe chemicals off the probe after cleaning.