Analog output configuration overview - GMP251

GMP251 User Guide

Document code
M211799EN
Revision
H
Language
English
Product
GMP251
Document type
User guide

The transmitter has 3 analog outputs that can be configured with the Insight PC software. Each output has the same set of configuration options in the analog output 1…3 menus under the Configure device menu (Analog output 1 shown in example).

The available selections depend on the probe that is connected to the transmitter (for example, CO2 output parameter selection is only available for CO2 probes).
Figure 1. Analog output configuration options example in Insight

The Analog output type selection in the General menu defines whether the outputs are used for voltage or current output. When you change the output type, Insight reloads the transmitter.

Figure 2. Analog output type (current or voltage) selection in Insight

Receiving analog output settings automatically from probe

The transmitter can receive the analog output settings automatically from the connected probe in the following cases:
  1. The transmitter is a new unit and has not been configured yet.
  2. The user has removed the analog output configuration from the transmitter.
It is recommended to always review the analog output settings with the Insight PC software after connecting a probe.