Other Commands - GMP251

GMP251 User Guide

Document code
M211799EN
Revision
G
Language
English (United States)
Product
GMP251
Document type
User guide
Table 1. Frestore Command
Syntax Description
frestore<cr> Restore the probe to its factory configuration. All user settings and user calibration parameters will be lost.
After using the frestore command, reset the probe using the reset command.
Example (restore the factory settings and reset the probe):
pass 1300
frestore
Parameters restored to factory defaults
reset
GMP251 1.0.0
Table 2. Pass Command
Syntax Description
pass [code]<cr> Access advanced serial commands.

Advanced commands can be used until the next reset.

code = Code for enabling advanced commands (1300).

Example:
pass 1300
Table 3. Reset Command
Syntax Description
reset<cr> Reset the probe. The probe will restart as if it had just been powered on.
Example:
reset
GMP251 1.0.0