Replacing the INTERCAP humidity sensor - GMW83 - GMW83A - GMW83D - GMW83DRP - GMW83DRPC - GMW83RPC - GMW84 - GMW84S - GMW86P - GMW87M

GMW80 Series User Guide

Document code
M211748EN
Revision
H
Language
English
Product
GMW83
GMW83A
GMW83D
GMW83DRP
GMW83DRPC
GMW83RPC
GMW84
GMW84S
GMW86P
GMW87M
Document type
User guide
  • INTERCAP® sensor (Vaisala item code 15778HM)
  • Flat head screwdriver (for opening the transmitter)
CAUTION To avoid contaminating or damaging the sensor:
  • Handle the sensor by its plastic frame.
  • Do not touch the sensor surface in the middle.
  • Wear lint-free ESD gloves.

To replace the sensor:

  1. Disconnect the transmitter body from the mounting base.
  2. Locate the INTERCAP® sensor on the component board. See Figure 2.
  3. Disconnect the old INTERCAP® sensor by pulling it straight out of its socket.
  4. Insert the new INTERCAP® sensor in the socket.
  5. Reconnect the transmitter to the mounting base.
  6. Check the output of the transmitter (or the display if included on the model) to verify that the humidity measurement is working normally and the transmitter is not in the error state. For more information on the error state, see Error state.