Connecting to Insight software - HMD110 - HMD112 - HMS110 - HMS112 - HMW110 - HMW112 - SWINSIGHT10 - TMD110 - TMI110 - TMW110

HMDW110 with TMI110 User Guide

Document code
M211726EN
Revision
F
Language
English
Product
HMD110
HMD112
HMS110
HMS112
HMW110
HMW112
TMD110
TMI110
TMW110
Document type
User guide
  • Computer with Vaisala Insight software installed
  • USB connection cable (item code 219690)
CAUTION

When connecting several devices at the same time, note that your computer may not be able to supply enough power through its USB ports. Use an externally powered USB hub that can supply >2 W for each port.

Before connecting the transmitter to a PC:

  • If the terminal block for terminals 5 ... 8 is wired, pull it out. This disconnects the transmitter from supply voltage and prevents possible equipment damage that may be caused by ground loops. For the locations of the terminal blocks and the service port, see the component board description.
  • If the other terminal block is wired (terminals 1 ... 4), pull it out also. This prevents the communication between the transmitter and the Modbus (RS-485) host from interfering with your connection.
  1. Open Insight software.
  2. Connect the USB cable to a free USB port on the PC.
  3. Connect the USB cable to the service port of the transmitter.
  4. Wait for Insight software to detect the transmitter.

    If the transmitter is not detected, disconnect and reconnect the Vaisala USB cable to the service port connector of the transmitter.