1hex |
Temperature measurement error. [44] |
2hex |
Humidity measurement error. [45] |
4hex |
Humidity sensor failure. [46] |
8hex |
Capacitance reference error. [47] |
10hex |
Ambient temperature out of range. [48] |
20hex |
Firmware checksum mismatch. [49] |
40hex |
Device settings corrupted. [50] |
80hex |
Additional configuration settings corrupted. [51] |
100hex |
Sensor coefficients corrupted. [52] |
200hex |
Main configuration settings corrupted. [53] |
800hex |
Supply voltage out of range. [55] |
2000hex |
Non-volatile memory read/write failure. [57] |
4000hex |
Calibration certificate checksum mismatch. [58] |