Test value registers - HMD110 - HMD112 - HMS110 - HMS112 - HMW110 - HMW112 - TMD110 - TMI110 - TMW110
HMDW110 with TMI110 User Guide
- Document code
- M211726EN
- Revision
- F
- Language
- English
- Product
- HMD110
- HMD112
- HMS110
- HMS112
- HMW110
- HMW112
- TMD110
- TMI110
- TMW110
- Document type
- User guide
Table 1. Modbus test registers
(read-only)Read the known test values from the test registers to verify the functionality of your Modbus implementation.
Register number (decimal) |
Address (hexadecimal) |
Register description |
Data format |
Test value |
7937 |
1F00hex |
Signed integer test |
16-bit integer |
-12345 |
7938 |
1F01hex |
Floating point test
|
32-bit float |
-123.45 |
7940 |
1F03hex |
Text string test |
7-character ASCII string |
Text string "-123.45" |