Syntax | Description |
---|---|
aerr ?<cr> | Show currently set analog output error level for both channels. |
aerr<cr> | Show currently set analog output error levels, prompt to enter new values. |
aerr [ch1] [ch2]<cr> |
Set error level for both channels without prompting. ch1 = error level for analog output channel 1 in mA. ch2 = error level for analog output channel 2 in mA. |
Example (set error level for both analog output channels to 21 mA):
|
Syntax | Description |
---|---|
amode<cr> | Show currently set analog output mode. |
Example:
|
Syntax | Description |
---|---|
aover<cr> | Show current aover status. |
aover [on|off]<cr> |
Set analog output overrange to: on = Allow output to exceed the scaled range by 10 %. off = Keep the analog output always in the scaled range. Note that error state behavior overrides this setting. |
Example (enable 10 % overrange for analog outputs):
|
Syntax | Description |
---|---|
asel ?<cr> | Show currently set analog output parameters and scaling. |
asel<cr> | Show currently set analog output parameters and scaling, prompt to enter new scaling values. |
asel [ch1 ch2]<cr> |
Set new output parameters for both channels, prompt to enter new scaling values. Selectable parameters for ch1 and ch2 are: t = temperature rh = relative humidity td = dew point h = enthalpy tw = wet bulb temperature |
asel [ch1 ch2 [ch1lo ch1hi ch2lo ch2hi]]<cr> |
Set analog output parameters and scaling for both channels without prompting. ch1 = Output parameter for channel 1. Parameters as above. ch2 = Output parameter for channel 2. Parameters as above. ch1lo = Low limit for channel 1 scaling. ch1hi = High limit for channel 1 scaling. ch2lo = Low limit for channel 2 scaling. ch2hi =High limit for channel 2 scaling. |
Example (show current analog output parameters and scaling):
|