Device information and status - HMD110 - HMD112 - HMS110 - HMS112 - HMW110 - HMW112 - TMD110 - TMI110 - TMW110

HMDW110 with TMI110 User Guide

Document code
M211726EN
Revision
F
Language
English
Product
HMD110
HMD112
HMS110
HMS112
HMW110
HMW112
TMD110
TMI110
TMW110
Document type
User guide
Table 1. ? command
Syntax Description
?<cr> Show listing of device information.
??<cr> Show listing of device information even if device is in poll mode and connection has not been opened using the open command.

Example:

?
HMDW110 / 1.0.6
Serial number  : H0134007
Batch number   : H0130002
Sensor number  : H0090003
Sensor model   : Humicap 180R
Order code     : HMD1102A1VA1
Cal. date      : 20140619
Cal. info      : VAISALA/HEL
Time           : 01:43:43
Serial mode    : STOP
Baud P D S     : 19200 N 8 1
Output interval: 5 S
Serial delay   : 25
Address        : 0
Filter         : 1.000
Ch1 output     : 4 ... 20 mA
Ch2 output     : 4 ... 20 mA
Ch1 RH lo      : 0.00 %RH
Ch1 RH hi      : 100.00 %RH
Ch2 T lo       : -40.00 'C
Ch2 T hi       : 60.00 'C
Table 2. Errs command
Syntax Description
errs<cr>

Show active error(s). Possible error messages are listed in Error messages in Insight software.

Example (no active errors):

errs
0000h
No errors
Table 3. Error messages on ERRS command
ERRS command response Corresponding error
T MEAS error Temperature measurement error. [44]
F meas error Humidity measurement error. [45]
RH sensor failure Humidity sensor failure. [46]
Frequency measurement outside the permissible value range Capacitance reference error. [47]
Ambient temperature error Ambient temperature out of range. [48]
Program flash check sum error Firmware checksum mismatch. [49]
Parameter flash check sum error Device settings corrupted. [50]
INFOA check sum error Additional configuration settings corrupted. [51]
SCOEFS check sum error Sensor coefficients corrupted. [52]
CURRENT check sum error Main configuration settings corrupted. [53]
Voltage error Supply voltage out of range. [55]
General flash failure w/r Non-volatile memory read/write failure. [57]
Calibration certificate check sum failure Calibration certificate checksum mismatch. [58]

Refer to Error messages in Insight software for more information about the error states.

Table 4. Help command
Syntax Description
help<cr> Show list of currently available serial commands.

Example:

help
??
?
ADDR
AERR
AMODE
AOVER
ASEL
ATEST
CDATE
CLOSE
CRH
CRHCLR
CT
CTCLR
CTEXT
ERRS
FCRH
FILT
FORM
FRESTORE
HELP
INTV
L
LI
OPEN
R
RESET
S
SDELAY
SEND
SERI
SMODE
SYSTEM
TIME
UNIT
Table 5. System command
Syntax Description
system<cr> Show firmware information.

Example:

system
Device Name : HMDW110
Copyright   : Copyright (c) Vaisala Oyj 2017. All
rights reserved.
SW Name     : HMP113
SW date     : 2018-01-03
SW version  : 2.2.3 HMDW
HW version  :
Table 6. Time command
Syntax Description
time<cr> Show transmitter uptime (time since last reset) in hh:mm:ss.

Example:

time
Time : 00:54:51