Table 1. Filt command
Syntax Description
filt [f.fff]<cr>

Set the speed at which the latest measurement result is integrated into the humidity and temperature readings. The command affects both analog output and serial line output.

f.fff = Measurement filtering setting, range 0.001 ... 1.0.

  • 1.0 = No filtering, latest measurement is output without averaging
  • 0.5 = Average of last two measurements
  • 0.1 = Average of approximately 16 measurements
filt<cr>

View the current setting. Also prompts you to enter a new value.

Example (view the current value and set filtering to value 0.5):

filt
Filter : 1.000 ? 0.5
Table 2. Frestore command
Syntax Description
frestore<cr>

Restores the factory default settings. All user-made settings are lost, including user calibration. Reset the transmitter after giving this command.

If you have replaced the HUMICAP® humidity sensor of the device yourself, you must redo the two-point humidity calibration using the FCRH command after performing a factory reset.

Example:

frestore
Factory settings restored
reset
HMDW110 / 2.2.3
Table 3. Reset command
Syntax Description
reset<cr> Resets the device.

Example:

reset
HMDW110 / 2.2.3