Syntax | Description |
---|---|
filt [f.fff]<cr> |
Set the speed at which the latest measurement result is integrated into the humidity and temperature readings. The command affects both analog output and serial line output. f.fff = Measurement filtering setting, range 0.001 ... 1.0.
|
filt<cr> |
View the current setting. Also prompts you to enter a new value. |
Example (view the current value and set filtering to value 0.5):
|
Syntax | Description |
---|---|
frestore<cr> |
Restores the factory default settings. All user-made settings are lost, including user calibration. Reset the transmitter after giving this command. If you have replaced the HUMICAP® humidity sensor of the device yourself, you must redo the two-point humidity calibration using the FCRH command after performing a factory reset. |
Example:
|
Syntax | Description |
---|---|
reset<cr> | Resets the device. |
Example:
|