Introduction to Vaisala Radiosonde RS41 - RS41-SGP - RS41-SG

RS41-SG and RS41-SGP User Guide

Product status
Document code
M211667EN
Revision
J
ft:locale
en-US
Product
RS41-SGP
RS41-SG
Document type
User guide

Vaisala Radiosonde RS41 offers excellent data availability and accuracy of humidity, temperature, pressure, and wind measurement. The radiosonde is fast and stable with individual, SI-standard traceable calibration.

Vaisala Radiosonde RS41 temperature sensor uses linear resistive platinum technology. The small size of the sensor results in low solar radiation error and guarantees fast response. Wind is derived from velocity based on satellite carrier frequency changes. Height and pressure are based on satellite ranging codes, either combined with differential corrections from the ground system, or using multiple GNSS systems which enhance accuracy and mitigate GNSS interference. Pressure calculation is supplemented with temperature and humidity data. Alternatively pressure is measured with a high-quality, shock-resistant capacitive silicon sensor which provides both height and pressure for meteorological messages.

Robust design, physical zero humidity check, and built-in functional temperature check ensure reliable performance in every situation. The radiosonde is also easy to use. For example, there is no need for the user to connect the batteries to the radiosonde to activate it. The radiosonde is automatically activated when placed on the ground check device. To make it easier to check the status of the radiosonde, the radiosonde has LED light indicators visible on the cover.

Vaisala Radiosonde RS41 with unwinder
1
Sensor boom
2
Unwinder pin
3
Power switch
4
Connector for additional sensor interface
5
Antenna
6
LED light
7
Unwinder
Operating environment
Property Description / Value
Function Radiosonde for atmospheric measurement, outdoor use

Maximum operating altitude

> 5000 m (16 404 ft), from ground level up to 40 km (131 234 ft)

Operating temperature

−90 … +60 °C (−130 …+140 °F)

Operating humidity

0–100 %RH

Power source Battery powered, no mains power
Overvoltage category CAT I

Use in wet location

Yes

Pollution degree

3