| Cable code | Cable length | Number of wires | PMU701 slot |
|---|---|---|---|
| 220497 | 10 m (32 ft 10 in) | 8 + shield 1 | G2 |
| 220498 | 30 m (98 ft 5 in) | ||
| 245931 | 50 m (164 ft 1 in) | ||
| 279350 | 100 m (328 ft 1 in) |
| Signal | Wire color | PMU701 connector pin | Wire color | Signal | |
|---|---|---|---|---|---|
| Not connected | – | 10 | 5 | Red | Ground |
| Not connected | – | 9 | 4 | Brown | RS-485 D− |
| Power in +24 V | Blue | 8 | 3 | Pink | RS-485 D+ |
| Not connected | – | 7 | 2 | – | Not connected |
| Not connected | – | 6 | 1 | – | Not connected |
| Color | Usage | Action |
|---|---|---|
| Shield | Ground |
Connect to PMU701 sensor grounding rail. |
| White | Not used (VOUT1) |
Cut to different lengths and tie together with electrical tape. |
| Green | Not used (AGND) | |
| Yellow | Not used (VOUT2) | |
| Gray | Not used (-) |
1 Calculated after cutting the cable.