DTS12G cable wiring - RWS200

RWS200 Installation Manual

Product status
Document code
M211608EN
Revision
AA
ft:locale
en-US
Product
RWS200
Document type
User guide

The DTS12G Type IIA cable is fixed to the sensor.

The cable can be lengthened by splicing it with a Type V extension cable.

DTS12G cables
Sensor code Cable length Number of wires DRI701 slot
DTS12G1 10 m (32 ft 10 in) 4 + shield 1

SENSOR

(first sensor on the right)

(second sensor on the left)

DTS12G2 20 m (65 ft 7 in)
DTS12G3 30 m (98 ft 5 in)
DTS12G5 50 m (164 ft 1 in)
DTS12G10 100 m (328 ft 1 in)
DTS12G15 150 m (492 ft 2 in)
DTS12G20 200 m (656 ft 2 in)
DTS12G cable termination to 8-pin connector

Signal

Sensor 2

Wire color

Sensor 2

DRI701 connector pin

Wire color

Sensor 1

Signal

Sensor 1

Analog ground Green 8 4 Green Analog ground
Base temperature sense − Red or brown 7 3 Red or brown Base temperature sense −
Base temperature sense + Blue or white 6 2 Blue or white Base temperature sense +
Excitation in Yellow 5 1 Yellow Excitation in
DTS12G cable shield and unused wires
Color Usage Action
Shield Ground

Connect to enclosure sensor grounding rail.

Gray Not used Cut.
DTS12G wiring
1 Calculated after cutting the cable.