| Cable code | Cable length | Number of wires | PMU701 slot |
|---|---|---|---|
| 223283 | 5 m (16 ft 9 in) | 8 + shield 1 | G2 |
| 220497 | 10 m (32 ft 10 in) |
| Signal | Wire color | PMU701 connector pin | Wire color | Signal | |
|---|---|---|---|---|---|
| Not connected | – | 10 | 5 | – | Not connected |
| Not connected | – | 9 | 4 | – | Not connected |
| Not connected | – | 8 | 3 | – | Not connected |
| Power in +12 V | Red | 7 | 2 | Gray | RS‑485 D− |
| Ground | Blue | 6 | 1 | Pink | RS‑485 D+ |
| Color | Usage | Action |
|---|---|---|
| Shield | Ground |
Connect to PMU701 sensor grounding rail. |
| White | Service port (RS‑232 ground) |
Cut to different lengths and tie together with electrical tape. |
| Green | Service port (RS‑232 RX output) | |
| Brown | Service port (RS‑232 RX input) | |
| Yellow | Not used | Cut and insulate with electrical tape. |
1 Calculated after cutting the
cable.