- Remove the screws holding the cover of the junction box. Remove the cover.
- Lead the sensor cable through the cable gland.
- Strip the cable and add cable ferrules to the wires.
-
For better protection against RF interference, fold the braided cable shield over the
plastic sleeve.
-
Connect the wires to the line terminal inside the junction box.
Signal Wire color WAC155 line terminal Heating 1 + Brown 1 Heating 2 − Gray 2 Not connected – 3 Vin + Pink 4 Vin − Green 5 Not connected – 6 SD1 − White 7 SD1 + Yellow 8 -
Tighten the output cable glands properly.
When installing the equipment for example in tropical, marine, or cold temperature environments, keep the equipment within its specific operating conditions. To avoid dust, dirt, or water to enter the equipment, tighten the cable glands properly. -
Insert the line terminal into the junction box and carefully reattach the cover.
Make sure that the gasket seals the junction box properly.