Table 1. PASS command
Syntax Description
PASS [code]<cr>

Access advanced serial commands. You can use advanced commands until the next reset.

[code] = Code for enabling advanced commands (1300).

Example:

pass 1300
Table 2. OPEN command
Syntax Description
OPEN [address]<cr>

Connect to a probe that is in POLL mode.

address = Probe address, range 0 ... 254.

Example (target probe in POLL mode, with address 52):

open 52
GMP231:  52 Opened for operator commands
Table 3. CLOSE command
Syntax Description
CLOSE<cr>

Close the connection to a probe.

Example (target probe in POLL mode, with address 52):

Example:

close
line closed
Table 4. ERRS command
Syntax Description
ERRS<cr>

Show active errors.

The possible errors and their identifying numbers correspond to content of the status word of the I2C interface.

Example (showing normal status, no active errors):

errs
NO CRITICAL ERRORS
NO ERRORS
NO WARNINGS
STATUS NORMAL

Example (showing probe in standby mode, no active errors):

errs
NO CRITICAL ERRORS
NO ERRORS
NO WARNINGS
STATUS 25            : Device in standby mode (set by user)
Table 5. RESET command
Syntax Description
RESET<cr> Reset the probe. The probe restarts as if it had just been powered on.

Example:

reset
GMP231 1.0.0
Table 6. FRESTORE command
Syntax Description
FRESTORE<cr>

Restore the probe to its factory configuration. All user settings are lost.

After using the FRESTORE command, reset the probe using the RESET command.

Example:

pass 1300
frestore
Parameters restored to factory defaults
reset
GMP231 - Version 1.1.0
Vaisala Oyj 2012 - 2014