Recommended installation - GMP231

GMP231 User Guide

Document code
M211501EN
Revision
E
Language
English
Product
GMP231
Document type
User guide

GMP231 is designed to be installed through a chamber wall, and attached to the chamber chassis using an attachment bracket and the two screw holes on the side of the probe. As the installation depth of the probe inside the chamber is critical, the mounting method must allow the probe to be positioned exactly.

The probe and electronics housing must remain outside the heated chamber. The electronics housing should be in a ventilated space that is open to ambient air.
Figure 1. Recommended installation
1
Ambient air
2
Chamber wall
3
Chamber interior
4
GMP231 electronics housing
5
M4 screw holes on both sides of the probe housing
6
8-pin M12 connector
7
Cable
8
CARBOCAP® sensor under PTFE filter. The filter must be inside the chamber completely.
9
Silicone plug. Insert the plug over the probe from inside the chamber.
10
Ø 44 mm installation tube through the chamber wall. The diameter is important for proper sealing by the silicone plug.