GMP231 sets the analog output channel into a defined error level instead of the measured result in 2 situations:

  • Probe detects a measurement malfunction. This means an actual measurement problem, such as sensor damage or unsuitable environmental conditions.
  • Measured value(s) are well outside the scaled output range.

The default error level depends on the output type:

  • For 4 ... 20 mA output, the default error level is 3.6 mA.
  • For 0 ... 20 mA output, the default error level is 23 mA.

The probe resumes normal operation of the analog output when the cause of the error state is removed.