Device information and status - GMP231

GMP231 User Guide

Document code
M211501EN
Revision
E
Language
English
Product
GMP231
Document type
User guide
Table 1. ? Command
Syntax Description
?<cr> Show listing of device information.
??<cr>

Show listing of device information even if the device is in POLL mode but its address is unknown.

Do not use the ?? command if you have more than one device on an RS-485 line.

Example:

?
Device              : GMP231
Copyright           : Copyright (c) Vaisala Oyj 2013. All rights reserved.
SW Name             : GMP231
SW version          : 1.0.1.1537
Snum                : J1320082
Calibrated          : 2014-03-11
Address             : 1
Smode               : STOP
Table 2. SNUM command
Syntax Description
SNUM<cr> Show serial number of the probe.

Example:

snum
Serial number       : J1320082
Table 3. VERS command
Syntax Description
VERS Show firmware version of the probe.

Example:

vers
GMP231 / 1.1.0.1537
Table 4. SYSTEM command
Syntax Description
SYSTEM<cr> Show probe firmware information.

Example:

system
Device Name         : GMP231
SW Name             : GMP231
SW version          : 1.1.0.1537
Table 5. TIME command
Syntax Description
TIME

Show how long the probe has been in operation since the last startup or reset.

The operation counter is in format hh:mm:ss. Full days and hours are retained in the counter, minutes and seconds are cleared at reset.

Example:

time
00000:23:56:49 Up time: 3409 s

Cumulative uptime is 23 hours, 56 minutes, and 49 seconds, uptime since last reset is 3409 seconds.

Table 6. HELP command
Syntax Description
HELP<cr> Show list of the currently available serial commands.

Example (shows basic commands, advanced commands are not enabled):

CLOSE
ECHO
ENV
ERRS
FORM
HELP
INTV
PASS
R
RESET
S
SDELAY
SEND
SERI
SMODE
SNUM
STANDBY
SYSTEM
TIME
VERS