When powered on, GMP231 starts up within 10 seconds. Measurements from the outputs (digital and analog) become available during this time but note that they only reach specified accuracy after a one minute warm-up period. For this reason, you should design the incubator’s control system so that it does not rely on measurements from GMP231 during this time.

Specifically note that the CO2 reading rises to the correct reading as the sensor's infrared emitter achieves operation temperature.