Property | Description/Value |
---|---|
Relative humidity | |
Measurement range | 0 … 100 %RH |
Accuracy: 1 | |
At temperature range 0 … +40 °C (+32 … +104 °F): | |
0 … 90 %RH | ±1.5 %RH |
90 … 100 %RH | ±2.5 %RH |
At temperature range −40 … 0 °C, +40 … +80 °C (−40 … +32 °F, +104 … +176 °F): | |
0 … 90 %RH | ±3.0 %RH |
90 … 100 %RH | ±4.0 %RH |
Factory calibration uncertainty at +20 °C (+68 °F): | |
0 … 90 %RH | ±1.1 %RH |
90 … 100 %RH | ±1.8 %RH |
Humidity sensor | HUMICAP® 180R |
Stability | ±2 %RH over 2 years |
Temperature | |
Measurement range | −40 … +80 °C (−40 … +176 °F) |
Accuracy over temperature range: | |
+15 … +25 °C (+59 … +77 °F) | ±0.2 °C (±0.36 °F) |
0 … +15 °C, +25 … +40 °C (+32 … 59 °F, +77 … +104 °F) | ±0.25 °C (±0.45 °F) |
−40 … 0 °C, +40 … +80 °C (−40 … +32 °F, +104 … +176 °F) | ±0.4 °C (±0.72 °F) |
Temperature sensor | Pt1000 RTD 1/3 Class B IEC 751 |
HMP110 probe | −40 … +80 °C (−40 … +176 °F) |
Storage temperature | −50 … +70 °C (−58 … +158 °F) |
Property | Description/Value |
---|---|
Resistive temperature input signals | |
Temperature sensor | Pt100 RTD / 4 wire, Class A IEC 751 |
Input impedance | 5.1K Ω |
Measurement range | −196 … +90 °C (−320.8 … +194 °F) |
Accuracy over temperature range: | |
−196 … −90 °C (−320.8 … −130 °F) −90 … −30 °C (−130 … −22 °F) −30 … 0 °C (−22 … +32 °F) 0 … +50 °C (+32 … +122 °F) +50 … +90 °C (+122 … +194 °F) |
±2.5 °C (±4.5 °F) ±0.75 °C (±1.35 °F) ±0.5 °C (±0.9 °F) ±0.25 °C (±0.45 °F) ±0.75 °C (±1.35 °F) |
Property | Description/Value |
---|---|
Current input signals | 0 … 22 mA |
Resolution | 0.67 μA |
Accuracy | ±0.15 % F.S. at +25 °C (+77 °F) |
Input impedance | 62 Ω |
Overload protection | 40 mA |
Voltage input signals | 0 … 5 V, 0 … 10 V |
Resolution | 0.0034 % F.S. |
Accuracy | ±0.15 % F.S. at +25 °C (+77 °F) |
Input impedance | 37K Ω |
Overload protection | 50 V max. |
Isolation | One common per logger |
Contact inputs | Open/Closed with magnetic reed switch cable connections (dry contact) |
Model | Measurement channels |
---|---|
HMT141 | 1 temperature and 1 humidity channel |
HMT143 | 2 RTD temperature channels |
HMT144 | 2 voltage inputs (0 … 5 VDC) |
HMT145 | 2 voltage inputs (0 … 10 VDC) |
HMT146 | 2 current inputs (0 … 20 mA) |
HMT147 | 2 door contacts |
HMT148 | 1 RTD temperature and 1 door contact |
HMT14D | 1 RTD temperature and 1 current input (0 … 20 mA) |
HMT14E | 1 RTD temperature and 1 voltage input (0 … 5 VDC) |
HMT14F | 1 RTD temperature and 1 voltage input (0 … 10 VDC) |
Property | Description/Value |
---|---|
Operating temperature | |
Data logger body, no display | −40 … +60 °C (−40 … +140 °F) |
Data logger body, with display | −20 … +60 °C (−4 … +140 °F) |
IP rating | IP65 |
Maximum operating altitude | 2000 m (6 500 ft) |
Property | Description/Value |
---|---|
Memory | 3,060 samples |
Sample rate | User selectable (interval: 2 … 60 minutes) |
Display (optional) |
128 × 64 resolution full graphics B&W display without backlight |
Property | Description/Value |
---|---|
Weight (with battery/without probe) | 300 g (10.6 oz) |
Screw terminals | 26 AWG … 20 AWG |
HMP110 probe interface | 4-pin M8 female panel connector |
HMP110 probe cable lengths | 3 m, 5 m, and 10 m (9.8 ft, 16.4 ft, and 32.8 ft) |
Hermetic door switch sensor cable length | 7.6 m (24.9 ft) |
Material | |
Data logger housing | PBT plastic |
Display window | PC plastic |
HMP110 probe body | Stainless steel (AISI 316) |
HMP110 probe grid filter | Chrome coated ABS plastic |
RTD temperature sensor | |
Sensor tip material | Stainless steel (AISI 316) |
Sensor tip length | 50.8 mm (2 in) |
Sensor tip diameter | 4.76 mm (0.19 in) |
Cable length | 5 m (16.4 ft) |
Property | Description/Value |
---|---|
Networking standards | IEEE 802.11b/g |
Data rates |
802.11b: up to 11 Mbps 802.11g: up to 54 Mbps |
Frequency band | 2402 ~ 2480 MHz |
Wi-Fi security | WEP (128-bit), WPA, WPA2 (Personal) |
Output power | +18 dBm (63 mW) |
Receiver sensitivity | −85 dBm typical |
Antenna | Onboard whip |
Contains | FCC ID: U3O-G2M5477 IC ID: 8169A-G2M5477 NCC ID: CCAF11LP0240T6 当該機器には電波法に基づく、技術基準適合証明等を受けた特定無線設備を装着している。 |
Property | Description/Value |
---|---|
EMC compatibility 2 |
EN 61326-1 EN 301 489-1 EN 300 328 V2.1.1 |
Electrical safety | EN 61010-1 |
Radio transmission equipment type approval | CMIIT ID: 2019DJ5109 |
Compliance marks | FCC, RCM, ROHS |
Region-specific standards and regulations are applied to wireless devices. The HMT140 logger can be used in the following regions/countries. The list is subject to change without prior notice.
|
Item | Order code |
---|---|
HMP110 | |
Humidity and temperature probe | HMP110 3 |
Humidity and temperature replacement probe | HMP110R 3 |
Humidity sensor | Vaisala HUMICAP® 180R |
Probe mounting flange | 226061 |
Probe mounting clamps, 10 pcs | 226067 |
Probe cable 3 m (9.8 ft) | HMT120Z300 |
Probe cable 5 m (16.4 ft) | HMT120Z500 |
Probe cable 10 m (32.8 ft) | HMT120Z1000 |
Duct installation kit | 215619 |
Sensor protection | |
Plastic grid filter | DRW010522SP |
Plastic grid with membrane filter | DRW010525SP |
Stainless steel sintered filter | HM46670SP |
Other accessories | |
RTD temperature probe 5 m (16.4 ft) | ASM210644SP |
Hermetic door switch sensor kit | 236319SP |
Thermal dampener blocks | 236310SP |
Four Dual Lock™ strips (76 mm/3 in) | 237217SP |
1 Incl. non-linearity, hysteresis,
and repeatability.
2 Due to updated regulations (ETSI EN
300 328 V2.2.2a), HMT140 will no longer be available for purchase in the following
countries from Aug. 6th, 2021, onwards: EU countries, Iceland, Liechtenstein,
Turkey, Norway, Switzerland, and the UK. The regulations do not affect devices
already in use.
3 See separate order form.