Connecting ozone sensor wires to OIF - RS41-SG - RS41-SGP

RS41 Ozone Soundings User Guide

Document code
M211486EN
Revision
G
ft:locale
en-US
Product
RS41-SG
RS41-SGP
Document type
User guide

The terminals on OIF are marked with a sticker attached on the card.

OIF terminals marked on sticker
OIF terminals
Number Connection Cable code
1 Radiosonde interface CBL210224
2 Add-on sensor IN for optional XDATA sensors -
3 Pump motor CBL210282
4 Ozone pump motor battery CBL210225
5 Heating battery CBL210295
6 Extra terminal -
7 SPC sensor W = white cable, B = blue cable -
Connect the ozone sensor wires to Ozone Interface Board OIF.
  1. Connect the white wire from the anode to the terminal on the corner of the interface W.
  2. Connect the blue wire from the cathode to the terminal B.