The heating battery is used in extreme
conditions. With RSA411, heating turns on automatically when the ozone pump temperature
drops under +5 °C, and turns off when the temperature rises above +7 °C. With RSA412,
heating turns on automatically when the ozone pump temperature drops under +9 °C, and turns
off when the temperature rises above +9.5 °C. There is no risk of overheating the box.
You must check the applicability of the heating on each site separately.
Connect the heating battery wire to the Heating battery terminal
on OIF.
Connect the red wire to the terminal +.
Connect the black wire to the terminal -.
Connecting heating battery
Attach 2-sided tape to the side and bottom of the battery to attach it to the sensor
frame.
2-sided tape attached to the
battery
Make sure to pass the battery wires between the ozone sensor frame wall and the ozone
sensor top cover and make sure that they are not twisted or under the sensor frame.
Battery wires running between the
ozone sensor frame wall and the ozone sensor