Pump efficiency correction (Cef) - RS41-SG - RS41-SGP

RS41 Ozone Soundings User Guide

Document code
M211486EN
Revision
G
ft:locale
en-US
Product
RS41-SG
RS41-SGP
Document type
User guide

The efficiency of the ECC-6A Ozonesonde air sampling pump decreases with altitude. Calculated ozone partial pressures must, therefore, be corrected for the efficiency loss. Correcting factors for ECC-6A pumps, with ECC sensor cathodes filled with 2.5 cm3 sensing solution and 3.0 cm3 sensing solution, are shown in the following table, respectively. At pressure level (P) the value of Cef is calculated using linear interpolation as a function of pressure.

The ASOPOS panel recommends the use of 3.0 cm3 only.
Ozone partial pressure correction factors
Ozone partial pressure correction factor Cef
Atmospheric pressure hPa Sensor cathode solution volume 2.5 cm3 Sensor cathode solution volume 3.0 cm3
2.0 1.160 1.171
3.0 1.124 1.131
5.0 1.087 1.092
10.0 1.054 1.055
20.0 1.033 1.032
30.0 1.024 1.022
50.0 1.015 1.015
100.0 1.010 1.011
200.0 1.007 1.008
300.0 1.005 1.006
500.0 1.002 1.004
1000.0 1.000 1.000

For other sensor manufacturers, the pump correction is very similar to the SPC ozone sensor, only the table values differ.

For Model Z, use the following table.

Ozone partial pressure correction factors
P/hPa Cef
≤3 1.24
5 1.124
7 1.087
10 1.066
15 1.048
20 1.041
30 1.029
50 1.018
70 1.013
100 1.007
150 1.002
≥200 1