Connecting Heating Battery to OIF411 (Optional) - RS41-D - RS41-SG - RS41-SGM - RS41-SGP

RS41 Ozone Soundings User Guide

Document code
M211486EN
Revision
E
Language
English
Product
RS41-D
RS41-SG
RS41-SGM
RS41-SGP
Document type
User guide

Cable needed: CBL210295

The heating battery is used in extreme conditions. Heating turns on automatically when the ozone pump temperature drops under +5 °C, and turns off when the temperature rises above +7 °C. There is no risk of overheating the box. However, the applicability of the heating should be checked on each site separately.

  1. Connect the heating battery wire to the Heating battery terminal on OIF411.
    1. Connect the red wire to the terminal marked with +.
    2. Connect the black wire to the terminal marked with -.
    Figure 1. Connecting Heating Battery
  2. Attach two-sided tape to the side and bottom of the battery to attach it to the sensor frame.
    Figure 2. Two-Sided Tape Attached to the Battery