The ozone destruction filter eliminates ozone generated during testing. Table 1 lists the parts required for the ozone destruction filter. However, the ASOPOS panel (Assessment of Standard Operating Procedures for Ozonesondes) recommends the use of purified, ozone-free air instead of the destruction filter. Refer to the GAW report 201 for more information. See Performance Review Literature .
Item Required | Vaisala Part Number |
---|---|
Particle and ozone filter: Mine Safety Appliances (MSA) Company. Delivered by MSA as a single cartridge (part number 815185). | 234561 (1 pc) |
Funnel, 75 mm in diameter, tube diameter 8 mm, glass | 12725 |
Connector tube 1: soft silicon tube approximately 5 cm long, inner diameter 6 mm, outer diameter 10 mm | 12642OS (set of tubes) |
Connector tube 2: soft vinyl tube approximately 60 cm long, I.D. 1/8" (3.2 mm), O.D. 1/4" (6.4 mm) | 12642OS (set of tubes) |
Connector tube 3: soft silicon tube approximately 2 cm long, inner diameter 2 mm, outer diameter 4 mm | 12642OS (set of tubes) |
Connector tube 4: Cut a piece of tubing from the ECC sensor air intake tube (2 cm), or order separately. | 17348S (by the meter) or SPC spare part No. OTU-19 |
Electrical tape (Nitto 15) | 4103 |
Assemble the ozone destruction filter components as shown in Figure 1.
- 1
- Catalyst
- 2
- Ultra fiber
- 3
- Funnel
- 4
- Connector tube 1
- 5
- Connector tube 2
- 6
- Connector tube 3
- 7
- Tape 1, 2, and 3 together with electrical tape