Use the FRESTORE command to restore the factory settings to the transmitter. All user settings will be lost, and the transmitter will return to the factory calibrated state (user adjustment is cleared).
FRESTORE<cr>
Example:
>frestore
Restoring factory defaults...done
Saving settings...done
>
CAUTION
If you have changed the HUMICAP® sensor yourself, be aware that the FRESTORE command clears all of the user adjustment parameters, including the capacitance offset and gain that are set by the CRH command. If you want to avoid having to perform the capacitance adjustment procedure again, record the adjustment parameters shown by the L command before performing the FRESTORE command. You can afterwards restore the adjustment parameters using the LI command. |
If the HUMICAP® sensor has been changed at a Vaisala Service Center, the factory calibration has been updated to match the new sensor, and it will not be affected by the FRESTORE command. |