Error Codes - DMT132

DMT132 User Guide

Document code
M211289EN
Revision
B
Language
English
Product
DMT132
Document type
User guide

The transmitter has error codes that can be displayed on the serial line using the ERRS command. See Display Currently Active Errors.

Table 1. Error Codes
Error Code Error Message Possible Cause(s)
E1 Temperature sensor open circuit/not calibrated
  • Temperature sensor damaged
  • Temperature sensor short circuit
  • Temperature sensor missing
E2 Humidity sensor open circuit/not calibrated
  • Condensation on humidity sensor
  • Humidity sensor damaged
  • Humidity sensor short circuit
  • Humidity sensor missing
E5 Capacitance out of range
  • Condensation on humidity sensor
  • Humidity sensor damaged
E6 Program checksum error Internal error
E7 Parameter storage checksum error Internal error
E9 Sensor coefficient checksum error Internal error
E10 Current settings checksum error Internal error
E11 Factory settings checksum error Internal error
E12 Temperature sensor values out of range
  • Temperature sensor damaged
  • Temperature sensor missing
E14 MeasTask not calculating (ATEST)

This error may occur when the analog output is in test mode.

See Test Analog Output.

E16 Default settings not found Internal error
E17 RH reading too high Internal error
E18 RH reading too low Internal error
E20 Log corrupted Internal error