Set LED Indicator Level - DMT132

DMT132 User Guide

Document code
M211289EN
Revision
B
Language
English
Product
DMT132
Document type
User guide

Use the ALARM command to change the dew point temperature above which the LED indicator plug is lit. The currently measured measurement parameter is used. You can also disable the LED indicator.

ALARM [on/off] [limit]<cr>
on/off
Enable (on) or disable (off) the LED indicator function.
limit
Dew point temperature above which the LED indicator is lit. Range -20 ... 50 °C.

Example (show the current status of the LED indication):

>alarm
Alarm Led           : OFF
Alarm level         : 20.000000
>

Example (enable the LED indicator at dew point above 20 °C):

>alarm on 20
Alarm level         : 20.000000
Saving settings...done
>