Terminal Application Settings - DMT132

DMT132 User Guide

Document code
M211289EN
Revision
B
Language
English
Product
DMT132
Document type
User guide

The serial interface settings of the DMT132 service port are presented in Table 1. The settings cannot be changed by the user.

Table 1. DMT132 Serial Interface Settings
Property Description / Value
Baud rate 19200
Parity None
Data bits 8
Stop bits 1
Flow control None

The steps below describe how to connect to the DMT132 using the PuTTY terminal application for Windows (available for download at www.vaisala.com/software) and a USB serial interface cable:

  1. Connect the USB serial interface cable between your PC and the RS-485 port of the DMT132 (connector II).
  2. Start the PuTTY application.
  3. Select the Serial settings category, and check that the correct COM port is selected in the Serial line to connect to field.
    You can check which port the USB cable is using with the Vaisala USB Instrument Finder program that has been installed in the Windows Start menu.
  4. Check that the other serial settings are correct for your connection, and change if necessary.

    Flow control should be set to None unless you have a reason to change it.

  5. Select Open to open the connection window and start using the serial line.
    If PuTTY is unable to open the serial port you selected, it will show you an error message instead. If this happens, restart PuTTY and check the settings.
  6. You may need to adjust the Local echo setting in the Terminal category to see what you are typing on the serial line. To access the configuration screen while a session is running, click the right mouse button over the session window, and select Change Settings... from the pop-up menu.
    Figure 1. PuTTY Terminal Application