Use the LI command to set the user adjustment parameters directly. This command is useful for restoring the desired adjustments, and removing the effects of user adjustment without having to perform the factory reset.
LI<cr>
After entering the command, you will be prompted to enter new values one at a time. Press ENTER without giving a value to move to the next parameter.
Example (shows default values):
>li
Cp offset : 0 ?
Cp gain : 1 ?
RH offset : 0 ?
RH gain : 1 ?
T offset : 0 ?
T gain : 1 ?
>