View User Adjustment Parameters - DMT132

DMT132 User Guide

Document code
M211289EN
Revision
B
Language
English
Product
DMT132
Document type
User guide

Use the L command to view the current user adjustment parameters that are affected by the CRH, RHCAL, CT, and LI commands. This command is useful for checking the currently applied correction.

L<cr>

The output values are as follows:

  • Cp offset and gain: capacitance, adjusted using the CRH command.
  • RH offset and gain: adjusted using the RHCAL command.
  • T offset and gain: adjusted using the CT command.

Example (shows default values, no user adjustment done):

>l
Cp offset           : 0
Cp gain             : 1
RH offset           : 0
RH gain             : 1
T offset            : 0
T gain              : 1
>