The transmitter has a built-in software function for testing analog outputs. To test the outputs:
  1. Make sure the transmitter is not in adjustment mode (see Calibration and adjustment overview).
  2. Press the + adjustment button (see Figure 1).
    This sets the output voltage level to the high point of the analog output range (for example 10 V). The output stays at this level for about 30 seconds after the + adjustment button has been pressed.
  3. Press the - adjustment button.
    This sets the output voltage level to the low point of the analog output range (for example 0 V). The output stays at this level for about 30 seconds after the - adjustment button has been pressed.
Figure 1. HMT130 output voltage measurement
If your transmitter has the optional display, the following texts are shown on the display the during analog output test:
  • Analog output test high: analog output set to upper limit
  • Analog output test low: analog output set to low limit