Measurement parameter configuration commands - HMT130

HMT130 User Guide

Document code
M211280EN
Revision
D
Language
English
Product
HMT130
Document type
User guide
Table 1. ENV command
Syntax Description
env [x.xxxx]<cr>
where
x.xxxx
Absolute pressure in the measuring environment. The default is 1.013. The pressure unit is bar.
Use the env command to show or set the environmental compensation values. In HMT130, the humidity measurement can be compensated for ambient pressure. The value set with the env command is a permanent value which remains in use after a restart.
Examples:
>env
Pressure (bar)      : 1.013
>
>env 0.980
Pressure (bar)      : 0.98
>save
Saving settings...done
>

You can use the Vaisala Humidity Calculator to simulate the effect of pressure change to dew point. The Humidity Calculator is available at:

www.vaisala.com/humiditycalculator