Check that the probes are located in equal conditions and wait until the readings have
stabilized (can take 30 minutes or more). If you are close to the probes, do not breathe
in their direction.
Press ADJUST to continue adjusting.
Choose To same as THI/II from the MI70 adjustment menu and press
SELECT. MI70 automatically recognizes which port the HMP70 series
probe is connected to.
Press YES to confirm the adjustment.
If the temperature difference between the reference probe and the
transmitter is too large, the adjustment cannot be done (MI70 notifies you of this). The
available reserve for T adjustment of the transmitter depends on the initial temperature
calibration of the unit.
The adjustment is done. Press BACK and
EXIT to return to the basic display.
Turn off the MI70 and detach the connection cable.