Accuracy of wet bulb temperature °C - HMT130

HMT130 User Guide

Document code
M211280EN
Revision
D
Language
English
Product
HMT130
Document type
User guide
Table 1. Accuracy of wet bulb temperature °C
Temp Relative humidity (%RH)
10 20 30 40 50 60 70 80 90 100
-40 °C 0.20 0.20 0.20 0.20 0.20 0.20 0.20 0.20
-20 °C 0.21 0.21 0.22 0.22 0.22 0.22 0.23 0.23
0 °C 0.27 0.28 0.28 0.29 0.29 0.29 0.30 0.30 0.31 0.31
20 °C 0.45 0.45 0.45 0.44 0.44 0.44 0.43 0.43 0.42 0.42
40 °C 0.84 0.77 0.72 0.67 0.64 0.61 0.58 0.56 0.54 0.52
60 °C 1.45 1.20 1.03 0.91 0.83 0.76 0.71 0.67 0.63 0.60
80 °C 2.23 1.64 1.32 1.13 0.99 0.89 0.82 0.76 0.72 0.68
100 °C 3.06 2.04 1.58 1.31 1.14 1.01 0.92 0.85 0.80 0.75
120 °C 3.85 2.40 1.81 1.48 1.28 1.13 1.03 0.95 0.88 0.83
140 °C 4.57 2.73 2.03 1.65 1.41 1.25 1.13 1.04 0.97 0.91
160 °C 5.25 3.06 2.25 1.82 1.55 1.37 1.24 1.13 1.05 0.99