Accuracy of wet bulb temperature °C - HMT130
HMT130 User Guide
- Document code
- M211280EN
- Revision
- D
- Language
- English
- Product
- HMT130
- Document type
- User guide
Table 1. Accuracy of wet bulb temperature
°C
Temp |
Relative humidity (%RH) |
10
|
20 |
30 |
40 |
50 |
60 |
70
|
80 |
90 |
100 |
-40 °C |
0.20 |
0.20 |
0.20 |
0.20 |
0.20 |
0.20 |
0.20 |
0.20 |
— |
— |
-20 °C |
0.21 |
0.21 |
0.22 |
0.22 |
0.22 |
0.22 |
0.23 |
0.23 |
— |
— |
0 °C |
0.27 |
0.28 |
0.28 |
0.29 |
0.29 |
0.29 |
0.30 |
0.30 |
0.31 |
0.31 |
20 °C |
0.45 |
0.45 |
0.45 |
0.44 |
0.44 |
0.44 |
0.43 |
0.43 |
0.42 |
0.42 |
40 °C |
0.84 |
0.77 |
0.72 |
0.67 |
0.64 |
0.61 |
0.58 |
0.56 |
0.54 |
0.52 |
60 °C |
1.45 |
1.20 |
1.03 |
0.91 |
0.83 |
0.76 |
0.71 |
0.67 |
0.63 |
0.60 |
80 °C |
2.23 |
1.64 |
1.32 |
1.13 |
0.99 |
0.89 |
0.82 |
0.76 |
0.72 |
0.68 |
100 °C |
3.06 |
2.04 |
1.58 |
1.31 |
1.14 |
1.01 |
0.92 |
0.85 |
0.80 |
0.75 |
120 °C |
3.85 |
2.40 |
1.81 |
1.48 |
1.28 |
1.13 |
1.03 |
0.95 |
0.88 |
0.83 |
140 °C |
4.57 |
2.73 |
2.03 |
1.65 |
1.41 |
1.25 |
1.13 |
1.04 |
0.97 |
0.91 |
160 °C |
5.25 |
3.06 |
2.25 |
1.82 |
1.55 |
1.37 |
1.24 |
1.13 |
1.05 |
0.99 |