Show Analog Output Status - HMM100

HMM100 User Guide

Document code
M211024EN
Revision
E
Language
English
Product
HMM100
Document type
User guide
STATUS [notification] [level]<cr>
notification
If specified, enables or disables the analog output error notification. Allowed values ON and OFF.
level
Error notification level. Allowed values are 0 … 30.

The STATUS command shows the following information:

  • Current analog output state (ON for normal, ERROR if an error is active)
  • Status of analog channel error notification (ON for enabled, OFF for disabled). See Analog Output Error Notification.
  • Scaling of the selected output quantity (low and high values)
  • Error notification level for the current analog output mode
Example (analog output in error state, output quantity RH, analog output mode 0 ... 10V):
>status
Channel 0:
Status              : ERROR
Notification        : ON
Level Lo            : 0
Level Hi            : 100
Error level         : 0.000000