Set Output Interval - HMM100

HMM100 User Guide

Document code
M211024EN
Revision
E
Language
English
Product
HMM100
Document type
User guide
INTV [n xxx]<cr>
n
Time interval in range 1 ... 255
xxx
time unit = "S", "MIN", or "H"

Use the INTV command to show or set the output interval of the serial line measurement messages (applies when the R command or RUN mode is used). The shortest output interval is one second. This command has no effect on the operation of the analog output.

Remember to save the changes!
Example:
>intv 1 s
Value               : 1
Unit                : S
>save
Saving settings...done
>