AMODE [mode]<cr>
- mode
- Analog output mode.
Use the AMODE command to show or set the operation mode of the analog output. This command is only needed with the 3-wire output module, as the output on the 2-wire model is always 4 ... 20 mA (loop powered).
If you change the analog output mode, you must calibrate the new mode using the ACAL command. See Calibrate Analog Output. |
Modes for 3-wire output model:
- 0..1V
- 0..5V
- 0..10V
- 0..20mA
- 4..20mA
Modes for 2-wire output model:
- 4..20mA
Remember to save the changes! |
Example (show current mode):
>amode
Ch1 : 0..10V
>save
Saving settings...done
>
Example (set mode to 0..20 mA):
>amode 0..20ma
Ch1 : 0..20MA
>save
Saving settings...done
>