Property | Description/Value |
---|---|
Relative Humidity | |
Measurement range | 0 … 100 %RH |
Factory calibration uncertainty (+20 °C / +68 °F) |
±1.5 %RH |
Humidity sensor |
HUMICAP® 180R |
Accuracy 1 | |
at −20 … +40 °C (−4 … +104 °F) |
±2 %RH (0 … 90 %RH) ±3 %RH (90 … 100 %RH) |
at −40 … −20 °C and +40 … +180 °C (−40 … −4 °F and +104 … +356 °F) |
±2.5 %RH (0 … 90 %RH) ±3.5 %RH (90 … 100 %RH) |
Dew Point Temperature | |
Measurement range | −20 … +100 °C (−4 … +212 °F) Td |
Accuracy 2 |
±2 °C (±3.6 °F) Td |
Property | Description/Value |
---|---|
EMC compliance |
Applicable parts of EN61326-1, generic environment |
Operating Temperature | |
Component board |
−5 … +55 °C (+23 … +131 °F) |
Stainless steel probe Plastic probe (standard body) Plastic probe (extended 400‑mm (16‑in) body) |
−70 … +180 °C (−94 … +356 °F) −70 … +180 °C (−94 … +356 °F) −70 … +120 °C (−94 … +248 °F) |
Plastic grid and membrane filter |
−20 … +80 °C (−4 … +176 °F) |
Porous PTFE, stainless steel sintered, and PPS plastic grid with stainless steel net filter |
−70 … +180 °C (−94 … +356 °F) |
Property | Description/Value |
---|---|
Power consumption |
6 mA (typical) |
Operating Voltage | |
2-wire model 3-wire model |
24 VDC 10 … 35 VDC or 24 VAC ±20 % 15 … 35 VDC or 24 VAC ±20 % when 0 … 10 V output is used |
Analog Output Types (1 Output Selectable) | |
2-wire model 3-wire model |
4 … 20 mA (loop-powered) 0 … 20 mA, 0 … 1/5/10 V |
Max. wire size | 0.5 … 1.5 mm2 (AWG) |
Service port | M8 connector for USB cable |
Output DC | Supply Voltage DC | Supply Voltage AC | Load Resistance | Typ. Current Consumption | Max. Current Consumption |
---|---|---|---|---|---|
0 ... 1 V | 10 ... 35 VDC | 24 VAC ±20 % | ≥ 10 kΩ | 6 mA | 10 mA |
0 ... 5 V | 10 ... 35 VDC | 24 VAC ±20 % | ≥ 10 kΩ | 6 mA | 10 mA |
0 ... 10 V | 15 ... 35 VDC | 24 VAC ±20 % | ≥ 10 kΩ | 6 mA | 10 mA |
0 ... 20 mA | 10 ... 35 VDC | 24 VAC ±20 % | 0 Ω | 6 ... 26 mA3 | 10 ... 30 mA3 |
0 ... 20 mA | 20 ... 35 VDC | 24 VAC ±20 % | 500 Ω | 6 ... 26 mA3 | 10 ... 30 mA3 |
Property | Description/Value |
---|---|
Service cable connector | M8 4-pin male |
Probe diameter | 12 mm (0.5 in) |
Probe Cable Lengths | |
Stainless steel probe Plastic probe (standard body) Plastic probe (extended 400‑mm (16-in) body) |
0.6/1.55/2.9 m (2/5.1/9.5 ft) 0.6/1.55/2.9 m (2/5.1/9.5 ft) 1.55 m (5.1 ft) |
Probe Material | |
Stainless steel probe Plastic probe Extension tube for 400‑mm probe |
AISI316/PPS PPS POM |
Probe mounting clamp | AISI316 |
Mounting Bracket Material | |
Lid |
ABS/PC |
Bottom plate |
Al |
Module housing material | ABS/PC (cover) |