L<cr>

Use the L command to view the current user adjustment parameters and trimmer adjustment. This command is useful for checking the currently applied correction.

The output values are as follows:

  • Cp offset and gain: capacitance, calibrated using the CRH command
  • RH trimmer offset and gain: correction by the trimmers
  • RH offset and gain: calibrated using the CRH command
  • T offset and gain: calibrated using the CT command
Example (shows default values, no user calibration done):
>l
Cp offset           : 0
Cp gain             : 1
RH offset trimmer   : 0
RH gain trimmer     : 1
RH offset           : 0
RH gain             : 1
T offset            : 0
T gain              : 1
>