Use the ACAL command to force the analog outputs to the following values:
- Current output: 2 mA and 18 mA
- Voltage output: 10 % and 90 % of the range
After entering the ACAL command, use a calibrated current/voltage meter to measure the output, and enter the values. Note that the calibration is only valid for the currently selected analog output modes.
| Syntax | Description |
|---|---|
| ACAL [channel]<cr> | [channel] = Analog output channel: 1 or 2 |
|
Example:
|
|