PTB330 specifications - PTB330

PTB330 User Guide

Product status
Document code
M210855EN
Revision
E
ft:locale
en-US
Product
PTB330
Document type
User guide
Measurement performance
Property Class A Class B
Barometric pressure range 500–1100 hPA
Linearity 1 ±0.05 hPa ±0.10 hPa
Hysteresis 1 ±0.03 hPa ±0.03 hPa
Repeatability 1 ±0.03 hPa ±0.03 hPa
Calibration uncertainty 2 ±0.07 hPa ±0.15 hPa
Accuracy at +20 °C (+68 °F) 3 ±0.10 hPa ±0.20 hPa
Barometric pressure range 50–1100 hPA
Linearity 1 ±0.20 hPa
Hysteresis 1 ±0.08 hPa
Repeatability 1 ±0.08 hPa
Calibration uncertainty 2 ±0.15 hPa
Accuracy at +20 °C (+68 °F) 3 ±0.20 hPa
Temperature dependence 4
500–1100 hPa ±0.1 hPa ±0.1 hPa
50–1100 hPa ±0.3 hPa ±0.3 hPa
Total accuracy −40 … +60 °C (−40 … +140 °F)
500–1100 hPa ±0.15 hPa ±0.25 hPa
50–1100 hPa ±0.45 hPa
Long-term stability
500–1100 hPa ±0.1 hPa/year ±0.1 hPa/year
50–1100 hPa ±0.2 hPa/year ±0.2 hPa/year
Miscellaneous
Pressure units hPa, mbar, kPa, Pa inHg, mmH20, mmHg, torr, psia hPa, mbar, kPa, Pa inHg, mmH20, mmHg, torr, psia
Resolution 0.01 hPa 0.1 hPa
Settling time at startup (1 sensor) 4 s 3 s
Response time (1 sensor) 2 s 1 s
Acceleration sensitivity Negligible
Maximum pressure limit 5000 hPa absolute
Maximum measurement rate 5 10 Hz
Inputs and outputs
Property Description/Value
Supply voltage 10–35 V DC
Supply voltage sensitivity Negligible
Typical power consumption at +20 °C (+68 °F) (voltage at 24 V DC with 1 pressure sensor)

RS‑232: 25 mA

RS‑485: 40 mA

Output voltage Uout: 25 mA

Output current Iout: 40 mA

Display and backlight: +20 mA

Serial communication RS-232, RS‑485, RS‑422
Operating environment
Property Description/Value

Operating pressure

500–1100 hPa, 50–1100 hPa

Operating temperature

PTB330 without display: −40 … +60 °C (−40 … +140 °F)

PTB330 with display: +0 … +60 °C (+32 … +140 °F)

Operating humidity

0–100 %RH, non‑condensing

Storage temperature

PTB330 without display: −55 … +80 °C (−67 … +176 °F)

PTB330 with display: −40 … +80 °C (−40 … +176 °F)

IP rating

IP66: Dust-tight. Protected from powerful water jets from any direction.

IP65 (NEMA 4) with local display: Dust-tight. Protected from water jets from any direction. 6

Mechanical specifications
Property Description/Value
Pressure fitting Barbed fitting for 1/8-inch (inside diameter) tubing or quick connector with shutoff valve for 1/8‑inch hose
Pressure connector M5 (10‑32) internal thread
Housing material G AlSi10 Mg (DIN 1725)
Weight 1–1.5 kg (2.2–3.3 lb)
Data transfer software
Property Description/Value

MI70 Link Interface software requirements

Microsoft® Windows OS

Microsoft® Excel

Analog output (optional)
Property Description/Value
Current output 0–20 mA, 4–20 mA
Voltage output 0–1 V, 0–5 V, 0–10 V
Accuracy at pressure range 500–1100 hPa 50–1100 hPa
At +20 °C (+68 °F) ±0.30 hPa ±0.40 hPa
At −40 … +60 °C (−40 … 140 °F) ±0.60 hPa ±0.75 hPa
Compliance
Property Value

EU directives and regulations

RoHS Directive (2011/65/EU) as amended by 2015/863

EMC Directive (2014/30/EU)

Low Voltage Directive (2014/35/EU), applies to units equipped with single-phase AC power supply Power-1

Electrical safety

EN 61010-1:2010 + A1:2019, applies to units equipped with single-phase AC power supply Power-1

Electromagnetic compatibility (EMC)

EN 61326‑1, industrial environment

EN 55011:2009 + A1:2010

Environmental

EN IEC 63000:2018

Accessories
Item Order code
Modules
Relay module RELAY-1L
Temperature-compensated analog output module AOUT-1T
Isolated RS-485 module RS485-1
Power supply module POWER-1
AC adapters for devices already equipped with an external AC adapter connector
AC adapter, EU MI70EUROADAPTER
AC adapter, USA MI70USADAPTER
AC adapter, UK MI70UKADAPTER
AC adapter, AUS MI70AUSADAPTER
Static pressure head
Static pressure head SPH10
Static pressure head with heating SPH20
Barometer mounting accessories
Junction box ASM211113
Wall mounting kit 214829
Installation kit for pole or pipeline 215108
Outdoor installation kit (weather shield) 215109
DIN rail clips with installation plate 215094
Panel mounting frame 216038
Connection cables
Connection cable for PTB330 and MI70 handheld meters for spot check or calibration and adjustment 211339
Service cables
USB-RJ45 serial connection cable 219685
D9-RJ45 serial connection cable 215005
Output cables for 8-pin connector
Connection cable 5 m with 8-pin M12 female, black 212142
Female connector 8-pin M12 with screw terminals 212416
Cable bushings
PTB330/220/PTU200 DC adapter and RS-232 cable for PC 213019
PTB330/PTB220/PTU200 DC adapter cable 213026
Others
Dust filter 237018SP
Barbed pressure fitting 1/8-inch 19498SP
Quick connector 1/8‑inch 220186
1 Defined as ±2 standard deviation limits of endpoint non-linearity, hysteresis, or repeatability error.
2 Defined as ±2 standard deviation limits of inaccuracy of the working standard including traceability to international standards.
3 Defined as the root sum of the squares (RSS) of endpoint non-linearity, hysteresis error, repeatability error, and calibration uncertainty at room temperature.
4 Defined as ±2 standard deviation limits of temperature dependence over the operating temperature range.
5 For class A you need a longer averaging time or measurement interval.
6 IP rating of PTB330AWS is IP40.