The INIE command resets the transmitter parameter set to factory default. Note that the current configuration settings are lost.

Before using this command, view the current settings with GETSET and save the output.

After using this command, you must reset the transmitter with the RESET command. This is to ensure that the default values are correctly validated.

The factory default values for all parameters are shown in the GETSET message example.

INIE <cr>

Example

>inie
FACTORY DEFAULTS
>