Grounding TERMBOX Signal Cable - TERMBOX

TERMBOX-9000, 1200, 1212, 1222 User Guide

Document code
M210310EN
Revision
G
Language
English
Product
TERMBOX
Document type
User guide

Ground the cable shield to the CG3 ... CG8 cable glands. Cable glands CG1 and CG2 cannot be grounded.

If cable glands CG3 ... CG8 are used, fold the cable shield back over the O‑ring of the plastic portion and mount the metal cable gland.

  1. Lead the sensor cable through the cable gland.
  2. Strip the cable and add cable ferrules to the wires.
  3. For better protection against RF interference, fold the braided cable shield over the plastic sleeve.
    Tighten the cable glands and make sure that no cable shield mesh gets on the circuit board.