Property | Description/Value |
---|---|
Response time (with probe and calibrator at the same temperature) |
With Vaisala sensors typically 10 min (deviation of the final value < ±1 %RH) |
Accuracy of thermometer | ±1 °C |
Accuracy of salt solutions | |
Lithium chloride LiCl | ±1.0 %RH + Greenspan's uncertainty 1 |
Magnesium chloride MgCl2 | ±1.0 %RH + Greenspan's uncertainty 1 |
Sodium chloride NaCl | ±1.4 %RH + Greenspan's uncertainty 1 |
Potassium chloride KCl | ±1.5 %RH + Greenspan's uncertainty 1 |
Potassium sulphate K2SO4 | ±1.5 %RH + Greenspan's uncertainty 1 |
Property | Description/Value |
---|---|
Operating temperature range | +0 … +50 °C (+32 … +122 °F) |
Property | Description/Value |
---|---|
Dimensions (H × W × L) | 90 × 230 × 200 mm (3.54 × 9.06 × 7.87 in) |
Weight | 1 kg (2.20 lb) without salt solutions |
Material (metal parts) | Anodized aluminum |
1 The uncertainty
given in Greenspan's calibration table at the calibration temperature. For
example, the accuracy of LiCl salt at +20 °C (+68 °F) is ±(1.0 + 0.3) %RH =
±1.3 %RH.