To get analog measurement data, an analog signal cable is needed (Vaisala item code 27168ZZ).
One scalable
voltage signal channel 0 … 1.0 V is available. You can scale the output to any output range
but it is recommended to have the scaling within the probe's measuring range to get accurate
measurements.
Connect the signal cable connector of the analog output to the base connector of the MI70 indicator. Connect the screw terminal block as follows:
Wire
Connect to
Brown wire
Common wire (−)
Yellow-green wire
Signal (+)
Open the menu by pressing
Open.
Select Functions and press .
Select Analog output and press .
To set the value for the 0.0 V output signal, select 0.0 V and press
Set. If the analog output is on, turn it off. Set the low value using the arrow buttons. To select the sign for the value, press +-. To confirm the setting, press
OK.
To set the value for the 1.0 V output signal, select 1.0 V and press
Set. Set the high value using the arrow buttons. To select the sign for the value, press +-. To confirm the setting, press
OK.
To activate the analog output and to return to the basic display, select Analog output on/off and
On. The wave icon appears in the upper left corner of the basic display.
To deactivate analog output, go to Menu > Functions > Analog output - Analog output on/off and press
Off.