Document type
User guide

The DC power is connected to a terminal stripe marked POWER in the lower right-hand corner of the motherboard. The 3 terminals are marked 31/+, 32/− and 33/(protective earth). The power terminal is directly connected to the exposed metal parts of the indicating transmitter DTR. Fuse voltage should be min. 48 V DC, max. size 4 A and fuse speed fast. Ensure that the protective fuse in the building system complies with local requirements.